would display the accumulated operating time of the blower (if the optional
active carbon filter is ordered and equipped, the LED screen would display the
accumulated operating time of the filter).
NOTE: The displayed figure needs to multiply by 10 to get the actual
operating time. The unit is hour.
d) Press the UP button to raise the front window to a proper height. Please
refer to 2.1.4 for the maximum opening of the front window
e) Press the FAN button to turn on the blower. The LED screen would
display the speed level of the fan memorized from the last time of operation. The
indicator light above the FAN button would be turned on to show the working
status of the blower. Make sure the blower runs at least FIVE minutes before
starting any experiment.
NOTE: The blower would be turned off automatically when the UV
lamp is turned on.
f) Press the LAMP button to turn on the fluorescent light. The indicator light
above the button would be turned on to show the working status of the
fluorescent light. Please refer to the actual condition of illumination in the
laboratory room to decide whether the fluorescent light is needed.
g) After finishing the experiment, turn off the blower and the fluorescent light and
press the DOWN button to close the front window
h) Press the UV button to turn on the UV light. The indicator light above the
button would be turned on to show the working status of the UV lamp. Please
make sure the sterilization is at least 30 minutes. Press the UV button again to
turn off the UV lamp. Please refer to the actual situation to decide whether
sterilization is needed.
NOTE: a) When the UV light is in working status, people should leave
the room in order to protect skin and eyes.