4. Assemble the calibration kit and install the calibration cap
on the device:
a. Insert the tab on the calibration cap into the left slot
on the device.
b. Press the calibration cap until it seats onto the
device.
c. Press both the top and right tab down onto the device
until they snap into place.
d. Ensure the calibration cap is properly seated.
e. Connect one end of the tubing to the calibration cap.
f. Connect the other end of the tubing to the gas
regulator (supplied in calibration kit).
5. Select Next.
6. Open the pressure reducer valve on the calibration gas cylinder.
7. The device will display the calibration results upon completion.
8. Close the valve after calibration.