MVE$Automation $Series$Quick$Reference$Guide
!
The following section describes how to adjust temperature alarm settings. At any time during the following procedure,
the user may exit the level by pressing the “Escape” button repeatedly, until the display returns to the “monitor” display
mode. After 30 seconds of inactivity, the controller will automatically return to the “monitor” display mode.
!
NOTE: Security Level 2 or higher is required to adjust temperature settings (see “Password and Security Setup”
section for details).
To exit any menu screen and return to the previous menu, press “ESC” key.
!
!
!
1. PRESS “SETUP”
The controller will prompt for a password. Use the “
▼”
keys to scroll to the appropriate number. Press “Enter” to move
cursor to the next position in the password.
!
!
!
2. PRESS “ENTER”
The display will read “Temperature Menus”.
!
!
!
!
3. PRESS “ENTER”
The display will read “Temp A Menu”.
NOTE: To adjust Temp B settings, press “Setup”
instead of “Enter”.
!
!
!
4. PRESS “ENTER”
The display will read “Temp A Enabled”.
NOTE: Probe A may be disabled now by pressing the
“
▼”
keys. Press “Enter” to save settings.
!
!
!
5. PRESS “SETUP”
Until Display Reads “Temp A High Alarm”. Use the “
▼
”
keys to adjust the temperature. Hold either button to scroll
quickly. Press “Enter” to save settings.
!
!
!
6. PRESS “SETUP”
The display will read “Temp A Low Alarm”. Use the “
▼”
keys to adjust the temperature. Hold either button to scroll
quickly. Press “Enter” to save settings
High
User
Level
Required
Use ▲ / ▼ To
Enter
Password
0000
!
!
Press Enter
For
Temperature
Menus
Or
Press Setup
For
Next
Menu
!
!
Press Enter
For
Temp
A Menu
Or
Press Setup
For
Next Menu
!
!
Temp
A
Enabled
Use
▲ / ▼ To Adjust
Press Enter
To Save
!
!
!
Temp A High
Alarm
Use ▲ /
▼
To Adjust
Press Enter
To Save
-110.0°C
!
!
!
!
Temp
A
Low
Alarm
Use ▲ /
▼
To Adjust
Press Enter
To save
-200.0°C
!
!
After all adjustments are complete, wait 30 seconds and the TEC 3000 will return to the Main Display screen
indicating new settings are in effect.