18
TROUBLESHOOTING
PROBLEM POSSIBLECAUSE SOLUTION
Thescreenshows
abnormalresult
Thecuposionwasnot
correctoritwasnotproperly
ghtened
Applythecucorrectlyand
tryagain.
Bodyposturewasnotcorrect
duringtesng
Reviewthe“BODYPOSTURE
DURINGMEASUREMENT”
seconsoftheinstrucons
andre-test.
Speaking,armorbody
movement,angry,excitedor
nervousduringtesng
Re-testwhencalmandwithout
speakingormovingduring
thetest.
Irregularheartbeat
(arrhythmia)
Itisinappropriateforpeople
withseriousarrhythmiatouse
thisElectronicSphygmoma
-
nometer.
Lowbaerysymbol LowBaery Changethebaeries
ThescreenshowsEr0
Pressuresystemisunstable
beforemeasurement
Don’tmoveandtryagain.
ThescreenshowsEr1 Failtodetectsystolicpressure
ThescreenshowsEr2
Failtodetectdiastolic
pressure
ThescreenshowsEr3
Pneumacsystemblocked
orcuistooghtduring
inaon
Applythecucorrectlyand
tryagain
ThescreenshowsEr4
Pneumacsystemleakage
orcuistoolooseduring
inaon
ThescreenshowsEr5
Cupressureabove
300mmHg
Switchthemonitoro,wait
30secondsandstartthe
procedurefromthestart.
ThescreenshowsEr6
Morethan3minuteswith
cupressureabove15mmHg
Measureagainaerve
minutes.Ifthemonitorissll
abnormal,pleasecontactthe
localdistributororthefactory.
ThescreenshowsEr7 Innermemoryerror
ThescreenshowsEr8
Deviceparameterchecking
error
ThescreenshowsErA
Pressuresensor
parameter error
PROBLEM POSSIBLECAUSE SOLUTION
Noresponsewhenyou
pressbuonorload
baery.
Incorrectoperaonorstrong
electromagnecinterference.
Removethebaeriesforve
minutes,andthenreinstallall
baeries.
MAINTENANCE
1. Donotdropthismonitororsubjectittostrongimpact.
2. Avoidhightemperatureandsolarisaon.Donotimmersethemonitorin
waterasthiswillresultindamagetothemonitor.
3. Ifthismonitorisstoredincoldtemperatures,allowittoacclimaseto
roomtemperaturebeforeuse.
4. Donotaempttodisassemblethismonitor.
5. Ifyoudonotusethemonitorforalongme,pleaseremovethe
baeries.
6. Itisrecommendedtheperformanceshouldbecheckedevery2yearsor
aerrepair.Pleasecontacttheservicecenter.
7. Cleanthemonitorwithadry,soclothorasoclothsqueezed
wellaermoistenedwithwater,diluteddisinfectantalcohol,ordiluted
detergent.
8. Nocomponentinsidethismonitorisserviceablebytheuser.Thecircuit
diagram,componentpartlist,calibraoninstruconsandother
informaonwhichcanassistanappropriatelyqualiedtechnicianto
repairthisdevice,canbesupplied.
9. Themonitorcanmaintainthesafetyandperformancecharacteriscsfor
aminimumof10,000measurementsorthreeyears,andthecu
integrityismaintainedaer1,000open–closecyclesoftheclosure..
10.Itisrecommendedthatthecushouldbedisinfected2mesevery
weekifneeded.Wipetheinnerside(thesidecontactsskin)ofthecu
withasoclothsqueezedaermoistenedwithEthylalcohol(75-90%),
thenallowittoairdry.
MX86005 - IFU - 11.2020.indd 18MX86005 - IFU - 11.2020.indd 18 2020/12/14 16:372020/12/14 16:37