18
19
TROUBLESHOOTING
PROBLEM POSSIBLECAUSE SOLUTION
Thescreenshows
abnormalresult
Thecuposionwasnot
correctoritwasnotproperly
ghtened
Applythecucorrectlyand
tryagain.
Bodyposturewasnotcorrect
duringtesng
Reviewthe“BODYPOSTURE
DURINGMEASUREMENT”
seconsoftheinstrucons
andre-test.
Speaking,armorbody
movement,angry,excitedor
nervousduringtesng
Re-testwhencalmandwithout
speakingormovingduring
thetest.
Irregularheartbeat
(arrhythmia)
Itisinappropriateforpeople
withseriousarrhythmiatouse
thisElectronicSphygmoma
-
nometer.
Lowbaerysymbol LowBaery Changethebaeries
ThescreenshowsEr0
Pressuresystemisunstable
beforemeasurement
Don’tmoveandtryagain.
ThescreenshowsEr1 Failtodetectsystolicpressure
ThescreenshowsEr2
Failtodetectdiastolic
pressure
ThescreenshowsEr3
Pneumacsystemblocked
orcuistooghtduring
inaon
Applythecucorrectlyand
tryagain
ThescreenshowsEr4
Pneumacsystemleakage
orcuistoolooseduring
inaon
ThescreenshowsEr5
Cupressureabove
300mmHg
Switchthemonitoro,wait
30secondsandstartthe
procedurefromthestart.
ThescreenshowsEr6
Morethan3minuteswith
cupressureabove15mmHg
Measureagainaerve
minutes.Ifthemonitorissll
abnormal,pleasecontactthe
localdistributororthefactory.
ThescreenshowsEr7 Innermemoryerror
ThescreenshowsEr8
Deviceparameterchecking
error
ThescreenshowsErA
Pressuresensor
parameter error
PROBLEM POSSIBLECAUSE SOLUTION
Noresponsewhenyou
pressbuonorload
baery.
Incorrectoperaonorstrong
electromagnecinterference.
Removethebaeriesforve
minutes,andthenreinstallall
baeries.
MAINTENANCE
1. Donotdropthismonitororsubjectittostrongimpact.
2. Avoidhightemperatureandsolarisaon.Donotimmersethemonitorin
waterasthiswillresultindamagetothemonitor.
3. Ifthismonitorisstoredincoldtemperatures,allowittoacclimaseto
roomtemperaturebeforeuse.
4. Donotaempttodisassemblethismonitor.
5. Ifyoudonotusethemonitorforalongme,pleaseremovethe
baeries.
6. Itisrecommendedtheperformanceshouldbecheckedevery2yearsor
aerrepair.Pleasecontacttheservicecenter.
7. Cleanthemonitorwithadry,soclothorasoclothsqueezed
wellaermoistenedwithwater,diluteddisinfectantalcohol,ordiluted
detergent.
8. Nocomponentinsidethismonitorisserviceablebytheuser.Thecircuit
diagram,componentpartlist,calibraoninstruconsandother
informaonwhichcanassistanappropriatelyqualiedtechnicianto
repairthisdevice,canbesupplied.
9. Themonitorcanmaintainthesafetyandperformancecharacteriscsfor
aminimumof10,000measurementsorthreeyears,andthecu
integrityismaintainedaer1,000open–closecyclesoftheclosure..
10.Itisrecommendedthatthecushouldbedisinfected2mesevery
weekifneeded.Wipetheinnerside(thesidecontactsskin)ofthecu
withasoclothsqueezedaermoistenedwithEthylalcohol(75-90%),
thenallowittoairdry.
MX86005 - IFU - 11.2020.indd 19MX86005 - IFU - 11.2020.indd 19 2020/12/14 16:372020/12/14 16:37