ADAMII™ © 2016 NanoEntek, Inc.
13
Home menu
Calibration
procedure
2. Shake bottle vigorously or vortex briefly for 5 seconds before use.
☞
Note:
ADAMII
TM
Calibration Beads is included in the kit. Calibration
Beads should be removed from the refrigerator and allowed to
room temperature before use.
3. Once again, vortex for 4 seconds and pipette for immediately.
4. Load the Calibration Beads into the ADAMII Assay Slide.
[Shake bottle vigorously]
[Vortexing]
5 sec.
4 sec.
[Vortexing]
[Loading volume: 25 uL]
[Pipetting Calibration Beads]