18
Home menu
Calibration
procedure
13. When the calibration is completed, the results will be displayed.
14. When you need previous results of calibration, click ‘Record’ button
of the ‘Setting’ menu.
Note: If any of the results are “FAIL”, please contact the manufacturer
or distributor.
Note: If any of the results are “Low”, please check the focus or
expiration date of the Calibration Beads. Or please re-test
using another slide.