2 Connection and commissioning
Narda FieldMan 27
EN
2.3 Power supply/rechargeable battery
operation
2.3.1 Operation with charger connected
⇒ Use a floating power supply for long-term operation, e.g. a suitable power bank
(PD 12 V, 2.5 A, BC1.2 and QC3.0).
2.3.2 Operation with rechargeable battery
⇒ Fully charge the rechargeable battery pack before commissioning.
Charging batteries
Starting the charging process:
1. Plug the included USB-C charging power supply with the matching country adapter
into a power outlet.
2. Connect the FieldMan with the supplied USB-C cable.
↳ The charging process begins.
2.4 Connect and test probe
Probes can be connected directly to the probe connector on the top of the FieldMan or to
the optical input:
• Connection at the probe connector: All digital HF and LFFieldMan probes
• Connection at the optical input: HP-01 and EHP-50F/G field analyzers.
Measuring with the charger connected can influence the measurement properties.
This means that compliance with the technical data is no longer guaranteed.
⇒ To ensure compliance with the permissible radio interference emissions, use only
the supplied charging power supply or the car charger (optional accessory) from
Narda.
Not all USB-C cables support fast charging with Power Delivery.
FieldMan_KBA.book Seite 27