EasyManua.ls Logo

National Instruments Data Acquisition Device NI USB-621x - Connecting PFI Input Signals; PFI Filters

National Instruments Data Acquisition Device NI USB-621x
185 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7 PFI
NI USB-621x User Manual 7-4 ni.com
Connecting PFI Input Signals
All PFI input connections are referenced to D GND. Figure 7-3 shows this
reference, and how to connect an external PFI 0 source and an external
PFI 2 source to two PFI terminals.
Figure 7-3. PFI Input Signals Connections
PFI Filters
You can enable a programmable debouncing filter on each PFI signal.
When the filters are enabled, your device samples the input on each rising
edge of a filter clock. USB-621x devices use an onboard oscillator to
generate the filter clock with a 40 MHz frequency.
Note NI-DAQmx only supports filters on counter inputs.
The following is an example of low to high transitions of the input signal.
High to low transitions work similarly.
PFI 0
Source
PFI 2
Source
USB-621x Device
D GND
PFI 2
PFI 0
I/O Connector

Table of Contents

Related product manuals