7-26 | ni.com
Chapter 7 Counters
Routing a Signal to Counter n Gate
Each counter has independent input selectors for the Counter n Gate signal. Any of the following
signals can be routed to the Counter n Gate input:
• RTSI <0..7>
• PFI <0..15>
• AI Reference Trigger (ai/ReferenceTrigger)
• AI Start Trigger (ai/StartTrigger)
• AI Sample Clock (ai/SampleClock)
• AI Convert Clock (ai/ConvertClock)
• AO Sample Clock (ao/SampleClock)
• DI Sample Clock (di/SampleClock)
• DO Sample Clock (do/SampleClock)
• PXI_STAR
• Change Detection Event
• Analog Comparison Event
In addition, Counter 1 Internal Output or Counter 1 Source can be routed to Counter 0 Gate.
Counter 0 Internal Output or Counter 0 Source can be routed to Counter 1 Gate.
Some of these options may not be available in some driver software.
Routing Counter n Gate to an Output Terminal
You can route Counter n Gate out to any PFI <0..15> or RTSI <0..7> terminal. All PFIs are set
to high-impedance at startup.
Counter n Aux Signal
The Counter n Aux signal indicates the first edge in a two-signal edge-separation measurement.
Routing a Signal to Counter n Aux
Each counter has independent input selectors for the Counter n Aux signal. Any of the following
signals can be routed to the Counter n Aux input:
• RTSI <0..7>
• PFI <0..15>
• AI Reference Trigger (ai/ReferenceTrigger)
• AI Start Trigger (ai/StartTrigger)
• PXI_STAR
• Analog Comparison Event
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments NI 6259 and is the answer not in the manual?

National Instruments NI 6259 Specifications

General IconGeneral
BrandNational Instruments
ModelNI 6259
CategoryData Loggers
LanguageEnglish

Related product manuals