M Series User Manual
Connecting Analog Input Signals
Table 4-3 summarizes the recommended input configuration for both types of signal sources.
Table 4-3. Analog Input Configuration
AI
Ground-Reference
Setting
*
Floating Signal Sources
(Not Connected to
Building Ground)
Ground-Referenced
Signal Sources
†
Examples:
• Ungrounded thermocouples
• Signal conditioning with isolated
outputs
• Battery devices
Example:
• Plug-in instruments with
non-isolated outputs
Differential (DIFF)
Non-Referenced
Single-Ended
(NRSE)
Referenced
Single-Ended (RSE)
*
Refer to the Analog Input Ground-Reference Settings section for descriptions of the RSE, NRSE, and
DIFF modes and software considerations.
†
Refer to the Connecting Ground-Referenced Signal Sources section for more information.
+
–
+
–
AI+
AI–
AI GND
Signal Source DAQ Device
+
–
+
–
AI+
AI–
AI GND
Signal Source DAQ Device
+
–
+
–
AI
AI SENSE
AI GND
Signal Source DAQ Device
+
–
+
–
AI
AI SENSE
AI GND
Signal Source DAQ Device
+
–
+
–
AI
AI GND
Signal Source DAQ Device
Ground-loop potential (V
A
– V
B
) are added
to measured signal.
NOT RECOMMENDED
+
–
+
–
AI
AI GND
V
B
V
A
Signal Source DAQ Device
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments NI 6259 and is the answer not in the manual?

National Instruments NI 6259 Specifications

General IconGeneral
BrandNational Instruments
ModelNI 6259
CategoryData Loggers
LanguageEnglish

Related product manuals