5-34 | ni.com
Chapter 5 Counters
In addition, TC or Gate from a counter can be routed to a different counter source.
Some of these options may not be available in some driver software. Refer to the Device Routing
in MAX topic in the NI-DAQmx Help or the LabVIEW Help for more information about available
routing options.
Routing Counter n Source to an Output Terminal
You can route Counter n Source out to any PFI terminal.
Counter n Gate Signal
The Counter n Gate signal can perform many different operations depending on the application
including starting and stopping the counter, and saving the counter contents.
Routing a Signal to Counter n Gate
Each counter has independent input selectors for the Counter n Gate signal. Any of the following
signals can be routed to the Counter n Gate input:
Any PFI terminal
AI Reference Trigger
AI Start Trigger
AO Sample Clock
DI Sample Clock
DI Reference Trigger
DO Sample Clock
Change Detection Event
Analog Comparison Event
In addition, a counters Internal Output or Source can be routed to a different counters gate.
Some of these options may not be available in some driver software. Refer to the Device Routing
in MAX topic in the NI-DAQmx Help or the LabVIEW Help for more information about available
routing options.
Routing Counter n Gate to an Output Terminal
You can route Counter n Gate out to any PFI terminal.