EasyManua.ls Logo

National Instruments NI PXIe-4139 - Page 15

National Instruments NI PXIe-4139
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 8. 100 µA to 100 mA Current Output and Measurement Verification
Level Range Limit
Range and
Limit
Test Point As-Found
Measurement Test
Limit (% of Current +
Offset)
As-Left Measurement
Test Limit (% of
Current + Offset)
100 µA 6 V -100 µA 0.022% + 2 nA 0.013% + 1.5 nA
100 µA
1 mA 6 V -1 mA 0.022% + 20 nA 0.013% + 15 nA
1 mA
10 mA 6 V -10 mA 0.022% + 200 nA 0.013% + 150 nA
10 mA
100 mA 6 V -100 mA 0.022% + 2 μA 0.013% + 1.5 μA
100 mA
1. Make the necessary connections for this procedure, as shown in the following figure:
Figure 8. Current Verification Connection Diagram
NI-DCPower Device
Sense LO
Sense HI
HI
LO
DMM
Current
Mode
+
2. Set the niDCPower Output Function property or NIDCPOWER_OUTPUT_FUNCTION
attribute to DC Current for the PXIe-4139.
3. Set the first specified level range, limit range, and limit on the PXIe-4139.
4. Measure the internal device temperature and perform self-calibration if necessary.
a) If the internal device temperature exceeds T
cal
±1 °C, wait up to five minutes for the
temperature to stabilize to within T
cal
±1 °C.
b) If after five minutes the stable temperature still exceeds T
cal
±1 °C, call the self-
calibration VI or function.
5. Set the level on the PXIe-4139 to the first specified test point.
NI PXIe-4139 Calibration Procedure | © National Instruments | 15

Related product manuals