EasyManua.ls Logo

National Instruments NI PXIe-4139 - Required Software; Related Documentation; Password; Calibration Interval

National Instruments NI PXIe-4139
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Required Software
Calibrating the PXIe-4139 requires you to install the following software on the calibration
system:
NI-DCPower 15.1 or later
Supported application development environment (ADE)—LabVIEW or
LabWindows
/CVI
Supported operating system—Windows
When you install NI-DCPower, you need to install support only for the application software
that you intend to use. Access calibration support in the locations shown in the following table:
ADE Calibration Support Location
LabVIEW NI-DCPower Calibration palette
LabWindows/CVI NI-DCPower function panel (niDCPower.fp)
You can download all required software from ni.com/downloads.
Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
NI PXIe-4138/4139 Getting Started Guide
NI DC Power Supplies and SMUs Help
PXIe-4139 Specifications
NI-DCPower Readme
LabVIEW Help
Visit ni.com/manuals for the latest versions of these documents.
Password
The default password for password-protected operations is NI.
Calibration Interval
Recommended calibration interval 1 year
2 | ni.com | NI PXIe-4139 Calibration Procedure

Related product manuals