NI PXI-4110 Calibration Procedure | © National Instruments | 21
Table 14. NI 4110 Voltage Line Regulation Verification Points
Channel V Range I Range I Limit V Level V
ext1
V
ext2
Measured Voltag e
Change
Limit (V)
Voltag e
Change (V)
V
1
(V) V
2
(V)
1 20 V 1A 100 mA 10 V 15.5 V 11 V
2 20 V 1A 100 mA -10 V 15.5 V 11 V