NI PXI-4110 Calibration Procedure | © National Instruments | 23
If the device has successfully passed all verification tests, the NI 4110 is within the published specifications, and adjustment is optional.
Table 16. NI 4110 Current Line Regulation Verification Points
Channel V Range I Range V Limit I Level V
ext1
V
ext2
Measured Current
Change
Limit (A)
Current
Change (A)
I
1
(A) I
2
(A)
1 20 V 20 mA 10 V 20 mA 15.5 V 11 V
1 20 V 1A 10 V 1A 15.5 V 11 V
2 20 V 20 mA 10 V -20 mA 15.5 V 11 V
2 20 V 1A 10 V -1 A 15.5 V 11 V