EasyManuals Logo

National Instruments PXI-4130 User Manual

National Instruments PXI-4130
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
NI PXI-4130 Calibration Procedure | © National Instruments | 41
1 200 mA 20 V -20 V 17 134 mA
18 156 mA
19 178 mA
20 200 mA
2A 20 V 1 20.0 mA
2 240 mA
3 460 mA
4 680 mA
5 900 mA
6 1.12 A
7 1.34 A
8 1.56 A
9 1.78 A
10 2.00 A
-20 V 11 20.0 mA
12 240 mA
13 460 mA
14 680 mA
15 900 mA
16 1.12 A
17 1.34 A
18 1.56 A
19 1.78 A
20 2.00 A
Table 20. NI 4130 Output Parameters and Test Limits for Current Programming
Accuracy Adjustment (Continued)
Channel
Current
Limit
Range
Voltag e
Level
Range
Voltag e
Level
Iteration Output