EasyManua.ls Logo

National Instruments PXIe-5160

National Instruments PXIe-5160
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Option Description
MAX Select the device in the Devices and Interfaces menu and click the Self-
Calibrate button.
When the two-year external calibration interval expires, an external calibration is required to
ensure performance within specifications over the subsequent two years.
First Measurements
Making a Measurement with InstrumentStudio
1. Connect CH 0 to an input signal.
2. Launch InstrumentStudio at Start»National Instruments.
3.
In the instrument header menu in the upper-right corner of the panel (
), add the
PXIe-5160/5162 to the large panel with Add/Remove Devices.
4. Click Auto to automatically configure device settings for the detected signal.
5. If the soft front panel is not already running, click Run/Stop.
6. Add oscilloscope measurements to the channel by selecting Add/Remove in the
measurement table of the oscilloscope.
For more information on the measurements available for oscilloscopes in InstrumentStudio,
refer to the InstrumentStudio Manual at ni.com/manuals.
Making a Measurement with LabVIEW
1. Launch LabVIEW.
2. Select Help»Find Examples.
3. Open the example VI that you want to use by selecting Hardware Input and Output»
Modular Instruments»NI-SCOPE (High-Speed Digitizers).
Tip If you are not sure which example to run, use the Quick Start VI, which is
found under Hardware Input and Output»Modular Instruments»NI-
SCOPE (High-Speed Digitizers)»Demos»niScope EX Quick Start.vi.
4. Follow any setup instructions in the VI and specify any desired settings.
5. Click Run to run the example program.
10 | ni.com | PXIe-5160/5162 Getting Started Guide

Related product manuals