EasyManua.ls Logo

National Instruments PXIe-5160 - Pxie-5160;5162 Self-Calibration; Running Self-Calibration

National Instruments PXIe-5160
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PXIe-5160/5162 Self-Calibration
To eliminate errors caused by changing temperatures, NI-SCOPE provides a highly repeatable
self-calibration function.
For the PXIe-5160/5162, self-calibration yields the following benefits:
Corrects for DC gain and offset errors within the digitizer by comparison to a precision,
high-stability internal voltage reference for all ranges, input impedance paths (50 Ω and
1 MΩ), and filter paths.
Calibrates trigger level offset and gain.
Calibrates trigger timing, as well as the time-to-digital conversion circuitry to ensure
accurate trigger timing and time-stamping.
Calibrates gain, offset, and phase for interleaved ADC modes, reducing ADC interleaving
errors in time-interleaved sampling modes.
Calibrates the phase DAC used to adjust the phase of the sample clock.
Self-calibration takes approximately two minutes to complete.
Running Self-Calibration
Running self-calibration enables the PXIe-5160/5162 to yield full performance over its
operating temperature range and two-year external calibration cycle.
Before running self-calibration:
Warm up the PXIe-5160/5162 for at least 15 minutes
When possible, disconnect the inputs
1
PXIe-5160/5162 modules are externally calibrated at the factory, but for optimal performance,
use self-calibration in any of the following situations:
When the module is placed in a new system
When the temperature changes more than ±3 °C from the previous self-calibration
Once 90 days have elapsed since the previous self-calibration
Note Unless temperature variations are a serious problem in your application, self-
calibration is not recommended more than once per day.
Tip You can use the nodes on the NI-SCOPE External Calibration palette to
programmatically return the date and module temperature of the previous self-
calibration. You can also view this information by selecting the device in MAX.
Self-calibrate the PXIe-5160/5162 programmatically in NI-SCOPE or on demand in MAX.
Option Description
NI-SCOPE Use the Self Calibrate node to self-calibrate the module.
1
If high-voltage, high-frequency signals are present during self-calibration, the calibration results
may be adversely affected or the calibration may fail with an error.
PXIe-5160/5162 Getting Started Guide | © National Instruments | 9

Related product manuals