Storage Environment
Ambient temperature range -40 °C to 71 °C (Tested in accordance
with IEC 60068-2-1 and IEC 60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)
Relative humidity range 5% to 95%, noncondensing (Tested in
accordance with IEC 60068-2-56.)
Shock and Vibration
Operating shock 30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC 60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)
Random vibration
Operating 5 Hz to 500 Hz, 0.3 g
rms
Nonoperating 5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC 60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)
Safety Voltages
Connect only voltages that are within these limits.
Supported measurement range -2 V to 6 V
Caution Do not connect the system to signals or use for measurements within
Measurement Categories II, III, or IV.
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
PXIe-6570 Specifications | © National Instruments | 15

Other manuals for National Instruments PXIe-6570

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXIe-6570 and is the answer not in the manual?

National Instruments PXIe-6570 Specifications

General IconGeneral
BrandNational Instruments
ModelPXIe-6570
CategoryIndustrial Electrical
LanguageEnglish

Related product manuals