RMX Programmable Power Supplies User Manual
Output Terminal Insulation
Caution Risk of electric shock. For safety reasons, even if the output terminal is 
grounded, secure reinforced or double insulation with a capacity greater than or equal 
to the isolation voltage of the RMX programmable power supply for the output 
terminals (including sensing terminals) and load cables. Refer to Appendix A, 
Specifications, for the isolation voltage of your model.
If cables with sufficient voltage rating cannot be obtained, use insulation tubes to 
secure reinforced or double insulation with a capacity greater than or equal to the 
isolation voltage of the RMX programmable power supply.
Confirm the voltage between any output terminal and ground is lower than the 
isolation voltage of the RMX power supply.
The signal cable may burn out. If the RMX programmable power supply is to be 
controlled through an external voltage (Vext), do not ground the external voltage 
(leave it floating).
The cable and load that are connected to the output terminal (including the sensing terminal) 
must have a reinforced or double insulation with a capacity that is greater than or equal to the 
isolation voltage of the RMX programmable power supply with respect to the chassis. Isolation 
voltage indicates the maximum allowed voltage that appears across the output terminal of the 
power supply unit and the protective conductor terminal (chassis terminal).
When the Output Terminal is Not Grounded (Floating)
The output terminal of the RMX programmable power supply is isolated from the protective 
conductor terminal. If you connect the GND wire of the power cord to the ground terminal of 
the switchboard, the chassis of the RMX programmable power supply is set to ground potential.
Pins 5 to 13 (for external control and parallel operation) and 18 to 25 (for external control and 
output monitoring) of the rear panel J1 connector are at approximately the same electric potential 
as the RMX programmable power supply negative output terminal. Cables and devices that are 
connected to these pins must have a reinforced or double insulation with a capacity greater than 
or equal to the isolation voltage of the RMX programmable power supply.