© National Instruments Corporation 3 NI USB-443x Calibration Procedure
Test Equipment
Table 1 lists the equipment National Instruments recommends for calibrating the NI USB-443x.
If the recommended equipment is not available, select a substitute calibration standard that meets the
given specifications.
Table 1. Recommended Equipment
Equipment Recommended Model
Parameter
Measured/
Adjusted
Minimum Requirements
Calibrator Fluke 5700A AI AC Coupled
Gain
AI DC Coupled
Gain
AI Offset
Frequency Range: 1 kHz
Voltage Range: up to 9 V
pk
AC Voltage Accuracy: ± 0.03% at
1kHz
*
DC Voltage Accuracy:
±150 ppm at 9 V
±100 μV with 1 mA load current
Function Generator HP/Agilent 33250A Timebase
Frequency
Accuracy
Frequency Range: up to 10 kHz
Frequency Accuracy: ±5 ppm
Voltage Range: up to 9 V
pk
Digital Multimeter NI 4071 AO Gain
AO Offset
DC Voltage Resolution: 1 μV
DC Voltage Accuracy at 3.15 V:
±150 ppm
BNC Shorting Cap
NI USB-4431:
quantity 4
NI USB-4432:
quantity 5
Pomona Electronics 5085 AI Offset Resistance: <1 Ω
BNC Cables:
quantity 6
Pomona Electronics 5697 All Characteristic Impedance: 50 Ω
BNC (Female) to
Banana Adapter
Pomona Electronics 1269 All
BNC T-Connectors:
quantity 4
Pomona Electronics 4896 AI Gain
AI Offset
Characteristic Impedance: 50 Ω
*
AC accuracy is the sum of all errors, including percent of reading error and temperature error, but excluding percent of
range error.
Frequency accuracy is the sum of all errors, including initial accuracy and stability errors.