User and Service Manual Natus® Brain Monitor & Embla Dx Series™
17
The current rating of the isolation transformer must be sufficient to operate all of the devices
powered by it. Refer to the current ratings of the isolation transformer and current rating for each
individual device connected.
Do NOT connect non-medical equipment which has been supplied as part of the system directly
to the wall outlet when the system is supplied, via MPSO, with a separating transformer.
Do NOT connect electrical equipment which has not been supplied as a part of the system to
the MPSO.
Ensure that the Brain Monitor & Embla Dx Series Amplifiers system is solely connected to a
three-wire, grounded, hospital-grade receptacle.
Patient Environment Warnings and Cautions
NOTE: The patient environment is defined as the area within 1.5 meters of the patient laterally
and within 2.5 meters of the floor in the area occupied by the patient.
Connect all patient electrodes to fully electrically isolated physiological devices only. Connecting
patient electrodes to any other device or external outlet may result in personal injury.
If a computer is located in the patient environment and is connected to a network, a network
isolator MUST be used.
The patient event button attached to the Brain Monitor & Embla Dx Series Amplifiers system is
NOT intended for critical patient-safety-related incidents.
Patient connections are NOT intended for direct cardiac contact.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
Do NOT touch any system accessible metal parts and the patient simultaneously.
Do NOT touch any earth-grounded components of the Brain Monitor & Embla Dx Series
Amplifier system and the patient simultaneously.
Do NOT allow loose electrodes to contact metal parts.
Do not use the Brain Monitor & Embla Dx Series Amplifier system in the vicinity of MRI or CT
systems.
Connection of a patient to high-frequency surgical equipment and to electroencephalography
equipment simultaneously may result in burns at the site of bio-potential input electrodes and
possible damage to the biological amplifiers. Please consult the user documentation of the
surgical equipment for instruction as to its proper use.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the natus neuro Brain Monitor and is the answer not in the manual?

natus neuro Brain Monitor Specifications

General IconGeneral
Brandnatus
Modelneuro Brain Monitor
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals