EasyManuals Logo

natus neuro Brain Monitor User Manual

natus neuro Brain Monitor
61 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
Natus® Brain Monitor & Embla Dx Series™ User and Service Manual
8
Safety & Standards Conformity
Standards of Compliance and Normative References
Multichannel Sleep/EEG Headbox System, Models: Brain Monitor, Embla NDx, Embla SDx Amplifiers,
detachable cord connected, portable 100-230Vac, 50/60Hz, 80VA.
1. Type of protection against electric shock: Class I
2. Degree of protection against electric shock: Type BF
3. Degree of protection against ingress of water: IPX0
4. Degree of safety of application in the presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
5. Mode of operation: Continuous
6. Environmental Conditions: Normal: 10-30°C, 30-75% rH, 700-1060hPa
The Brain Monitor & Embla Dx Series Amplifiers and accessories have been designed to comply with
the following national and international standards.
Safety Standard of Compliance and Normative References
IEC 60601-1:2012 - General Safety Third Edition
CAN / CSA-C22.2 No. 60601-1: 08(R2013) +C2:2011
IEC 60601-1-6:2010 Usability Third Edition
IEC 62366:2007, Edition 1.0
IEC 60601-2-26:2012 Electroencephalographs Third Edition
IEC60601-2-61:2011 - Pulse Oximeters
EN ISO 80601-2-61:2011, Edition 1
EMC Standard of Compliance and Normative References
IEC 60601-1-2:2014 EMC Fourth Edition
IEC 61000-3-2:2014, Fourth Edition
IEC 61000-3-2 Harmonic emissions Class A
IEC 61000-3-3:2013, Third Edition Voltage Fluctuations/ Flicker emissions
CISPR11, Edition 5.0 A1:2010 RF emissions Group 1, Class A
IEC 61000-4-2:2008, Second Edition
IEC 61000-4-2 Electrostatic Discharge (ESD) ±8 kV contact (to the patient lead and expose
metal), ±15 kV air
IEC 61000-4-3 Third Edition with A1:2007+A2:2010
IEC 61000-4-3, 3 Vrms, 80 MHz to 2.7 GHz
IEC 61000-4-4:2012, Third Edition, fast transient/burst ±2 kV power supply ±1 kV
IEC 61000-4-5:2014, Third Edition
IEC 61000-4-5 ±1 kV Surge differential mode ±2 kV common mode
IEC 61000-4-6 Second Edition with A1:2004 + A2:2006
IEC 61000-4-6, 150 kHz to 80 MHz
IEC 61000-4-8:2009, Second Edition
IEC 61000-4-8, Power frequency (50/60 Hz) magnetic field, 30 A/m

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the natus neuro Brain Monitor and is the answer not in the manual?

natus neuro Brain Monitor Specifications

General IconGeneral
Brandnatus
Modelneuro Brain Monitor
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals