6.3 Storing
After use, always store the pH meter disinfected, cleaned and with the protective cap
attached (see chapter 6.2 Disinfection) in the supplied NAWA bag.
Before storing, always spray a pump of NAWAlution into the protective cap so
that it can be easily removed before the pH meter's next use!
Storage at temperatures below 10°C and above 40°C reduces the service life of the pH
meter (see Chapter 7 Technical data).
6.4 Error indications
No display
acticvity
Check if the meter is sufficiently charged.
Charge the meter for about 1 hour.
If this error persists, contact the NAWA Technology
Ltd. customer service.
Individual steps
cannot be
confirmed with
the touch button
The touch sensor responds to tapping, not pressing.
Lightly tap the button.
If this error persists, contact the NAWA Technology
Ltd. customer service.
pH values and /
or temperature
are out of the
specified
measuring range
Make sure to moisten the sensor in the probe tip before
each use or after longer inactivity (from 14 days).
Recalibrate the device.
If this error persists, contact the NAWA Technology
Ltd. customer service.
The device skips
a step during
calibration
Finish the calibration as specified.
Recalibrate the device.
If this error persists, contact the NAWA Technology
Ltd. customer service.
The display
shows "PROBE
LOST"
Check the connection between the probe and the
measuring device.
Disconnect the probe from the measuring device.
Reconnect the probe and the measuring device via
bayonet lock.
If this error persists, contact the NAWA Technology
Ltd. customer service.