Version 13 – 01/2019
8
Place the probe (4) (without protective cap) in the test tube (1) for approx. 15
min.
Then carefully dab the probe tip with sterile gauze.
4.2.4 Setting the time
Set the current date while preparing the ISFET-sensor in the probe tip (point 4.2.3).
The meter takes over the time zone of your PC (Windows and iOS)
Connect the measuring device to the charging station (bayonet lock).
Connect charging station (8) and USB cable via USB port to
PC/laptop/tablet PC.
Open the folder of the storage medium.
Double-click the "createsystemtime.exe" file (located on the meter's memory
and on the included USB stick).
Safely remove the USB connection ("eject" the meter).
Disconnect the meter (3) from the charging station (8) (open the bayonet lock).
4.3 Activating and deactivating the device
Activation:
The meter turns on automatically as soon as probe and measuring device are
connected (bayonet lock).
While measuring, all data is automatically stored on the meter and can be accessed
via PC/laptop/tablet PC.
Confirming actions:
During the calibration or measuring process, the display prompts you to confirm
individual steps: Tap the touch button with your thumb or forefinger, do not press.
Deactivation:
The display will blank out, if probe and measuring device are disconnected (bayonet
lock).
The data remains stored on the meter until manually cleared.