EasyManua.ls Logo

Neen PERICALM - Continence Treatment Programs

Neen PERICALM
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PERICALM™ Instruction Manual - GB
Continence Treatment Programs
The parameters in the table are suggested parameters only and should be used as that.
Please refer to the ACA Notes on Good Practice or The Chartered Society of
Physiotherapy Guidelines, to view the most up to date neuromuscular electrical
stimulation parameters.
As a clinician, you may prefer to use your own parameters to treat patients and it is
then ideal for you to use the customisable programs MEM1-3.
11
6 Sequential phases:
Phase 1: 2Hz for 4 min,
Phase 2: 10Hz for 10 min,
Phase 3: 15Hz for 5 min,
Phase 4: 20Hz for 5 min,
Phase 5: 30Hz for 5 min,
Phase 6: 10Hz for 5 min
All phases are 5 seconds work, 5 seconds rest at 220 µS,
0.8 seconds ramp.
Programs Hz PW
Ramp up Work Rest Overall
Rate μS
time in time in time in time in
seconds seconds seconds seconds
PAIN RELIEF 2 175 1 Cont Cont 25 min
PR
STRESS 1 30 200 1 5 5 25 min
STR1
STRESS 2 35 210 1 5 10 25 min
STR2
URGE 10 210 1 5 6 25 min
URGE
FREQUENCY/URGE
10 250 1 5 10 25 min
FURG
PELVIC FLOOR
WORK OUT AND
LACK OF
SENSITIVITY
(34 MIN)
PFW
CUSTOMISED
2 - 100 50 - 450
0.3 - 9.9 2 - 99 2 - 99 5 - 60
MEM 1
CUSTOMISED
2 - 100 50 - 450
0.3 - 9.9 2 - 99 2 - 99 5 - 60
MEM 2
CUSTOMISED
2 - 100 50 - 450 0.3 - 9.9
2 - 99 2 - 99 5 - 60
MEM 3

Related product manuals