PERICALM™ Instruction Manual - GB
4
Warnings
This unit must be used with the guidance of a Physiotherapist,
Doctor or Continence Advisor
Type BF equipment, Continuous Operation
Do not insert lead wires into a mains power supply
Do not immerse the unit into water or any other substance
Do not use the unit in the presence of a flammable anaesthetic gas
mixture and air or with oxygen or nitrous oxide
If using rechargeable 9 Volt PP3 Nickel Metal Hydride batteries,
be sure to use a CE approved battery charger. Never connect the
PERICALM™ Continence directly to a battery charger or to any other
mains powered equipment.
Never connect the PERICALM™ directly to a battery charger or any other
mains powered equipment
Patient Electrodes are for single patient use only
Use only CE approved vaginal or anal electrodes i.e. PERIFORM
®
+
or ANUFORM
®
Keep out of reach of children
To avoid the effects of electromagnetic interference never use
the PERICALM™ device, within 4 meters of a cellular telephone or
near any other powerful radio interference producing equipment that
produces electrical sparks etc. In the electromyography (EMG) mode,
the PERICALM™ device may be susceptible to strong interfering radio
type emissions that may lead to temporarily increased EMG microvolt
readings. The reading will immediately return to the correct value when
the interference ceases. (Remember that a relaxed muscle should read
below 3.5 µvolts)