PERICALM™ Instruction Manual - GB
Specifications
17
1. Dual channel: individually isolated circuits.
2. Amplitude: 0 - 80 mA into 500 Ohm load; indication only. Actual mA
will tend to be less than indicated due to electrode impedance: at
1000 Ohms load (Electrodes in poor condition) the maximum will be
limited to 70 mA, at 1500 Ohms load the maximum will be limited to
65 mA.
3. Type: Constant Current, maximum output voltage 180 Volts
+10 / -30 Volts
4. Waveform: Asymmetrical, rectangular bi-phasic with zero
DC current.
5. Selectable pulse width: 50 µS – 450 µS (2% accuracy)
6. Pulse Rate selection: in the continuous mode 2 – 100 Hz
(2% accuracy)
7. Time duration of the treatment: 5 - 60 minutes.
(Custom programs only)
8.
Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2
volts the
battery symbol will flash on/off once every second.
9. Open Electrode Detect: If an open circuit is detected at the output
of channel A or B by: a probe not being attached, or with a probe
attached but not in position, when the unit intensity is ramped up to
5mA the output current will be reset at zero.
10. Ramp up time 0.3 - 9.9 seconds.
11. If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not
turn on.
12. Physical dimensions: 80 x 67 x 45 mm.
13. Weight: 90 grams without battery, 125 grams with battery.
14. Environmental conditions for storage & transport: -10 to +50 degrees
Centigrade 0-90% Humidity.