PERICALM™ Instruction Manual - GB
20
150 kHz to
80 MHz
80 MHz to
800 MHz
800 MHz to 2,5
GHz
d =1.2 Pd =1.2 Pd = 2.3 P
0,01 0.12 0.12 0.23
0,10.380.3
80
.73
11.2 1.
22
.3
10 3.83.8 7.3
100 12 12 23
and mobile RF communications equipment and PERICALM™
Table 206: Recommended separation distances between portable
product
The PERICALM™ product is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the
PERICALM™ product can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the PERICALM™ product as
recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of
transmitter
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters [m] can be estimated using the
equation applicable to the frequency of the transmitter, wh ere P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.