MCS260 
CORNERSTONE 260 MONOCHROMATORS 
18 
 
 
 
5.2  MICROMETER ADJUSTABLE SLITS 
 
Micrometer adjustable slit assemblies are continuously variable from fully closed to 3 mm width. A 
height adjustment slide allows variation in the height from 2 to 12 mm. Benefits of the micrometer 
adjustable slits are flexibility and high throughput. This type of slit is designed primarily for 
versatility and convenience in changing resolution and throughput, which are related to the slit 
width.  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 9:  A Micrometer Adjustable Slit 
The slit width setting is read on the micrometer.  A set of numbers go around the turning dial.  
Another set of numbers are located on the shaft.  When the zeroes in both these locations line up, 
the slit is fully closed.  Turning the dial clockwise advances the dial position further down on the 
shaft, closer to the body of the micrometer.  This opens the slit.   
 
Use a 10x multiplier to convert the micrometer reading to the actual slit opening size.  For 
example, turning the dial one full revolution starting from the fully closed position will give a 
reading of 50 on the micrometer.  Using the multiplier, this indicates the micrometer width is set to 
500 um. If unsure of the reading, begin at the fully closed position and add up each full revolution 
made. 
 
Figure 10:  A Fully Closed Micrometer Adjustable Slit 
 
The micrometer is used to make the slit 
The slide is used to