EasyManua.ls Logo

NI 9229 - Test Equipment; Test Conditions; Initial Setup

NI 9229
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NI 9229/9239 Calibration Procedure | © National Instruments | 3
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
Test Conditions
The following setup and environmental conditions are required to ensure the NI 9229/9239
meets published specifications.
Keep connections to the device as short as possible. Long cables and wires act as antennas,
picking up extra noise that can affect measurements.
Verify that all connections to the NI 9229/9239 are secure.
(NI 9229/9239 with screw-terminal) Use shielded copper wire for all cable connections
to the device.
Use twisted-pairs wire to eliminate noise and thermal offsets.
Maintain an ambient temperature of 23 °C ± 5 °C.
Keep relative humidity below 80%.
Allow a warm-up time of at least 10 minutes to ensure that the NI 9229/9239 measurement
circuitry is at a stable operating temperature.
Initial Setup
Complete the following steps to set up the NI 9229/9239.
1. Install the NI-DAQmx driver software.
2. Make sure the NI cDAQ-9178 power source is not connected.
Table 1. Recommended Equipment
Equipment
Recommended
Model
Parameter
Measured
Requirements
Calibrator Fluke 5700A Accuracy,
Adjustment
Use a high-precision voltage source
with an accuracy 100 ppm and an
output impedance 50 .
Function
Generator
NI 4461 Gain Matching,
Phase Matching
Use a function generator capable of
sourcing both 1 kHz and 20 kHz
±10 V
pk
AC sinusoidal signals with
amplitude uncertainty of ±10%.
Chassis NI cDAQ-9178 All

Related product manuals