EasyManua.ls Logo

NI 9229 - Verification Procedures; Accuracy Verification

NI 9229
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 | ni.com | NI 9229/9239 Calibration Procedure
3. Connect the NI cDAQ-9178 to the system safety ground.
a. Attach a ring lug to a 14 AWG (1.6 mm) wire.
b. Connect the ring lug to the ground terminal on the side of the NI cDAQ-9178 using
the ground screw.
c. Attach the other end of the wire to the system safety ground.
4. Install the NI 9229/9239 in slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of
the NI cDAQ-9178 chassis empty.
5. Connect the NI cDAQ-9178 chassis to your host computer.
6. Connect the power source to the NI cDAQ-9178 chassis.
7. Launch Measurements & Automation Explorer (MAX).
8. Right-click the device name and select Self-Test to ensure that the device is working
properly.
Verification
The following performance verification procedures describe the sequence of operation and
provide test points required to verify the NI 9229/9239. The verification procedures assume that
adequate traceable uncertainties are available for the calibration references.
Accuracy Verification
Complete the following procedure to determine the As-Found status of the NI 9229/9239.
1. Set the calibrator to Standby mode (STBY).
2. Connect the NI 9229/9239 to the calibrator as shown in Figure 1.
Figure 1. Accuracy Verification Connections
3. Set the calibrator voltage to a Test Point value indicated in Tables 4 or 5.
4. Set the calibrator to Operate mode (OPR).
NI 9229/9239 with
Screw-Terminal
Calibrator
HI
LO
AI1
+
AI0
+
AI2
+
AI3
+
NI 9229/9239
with BNC
Calibrator
HI
LO
AI1
AI0
AI2
AI3
+
+
+
+

Related product manuals