EasyManua.ls Logo

Nikon SMZ1000 - Lighting Adjustment and P-ICI2 Illuminator

Nikon SMZ1000
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
5
Lighting area adjustment [G-LS]
It is possible to adjust the lighting area by moving the lamp socket back
and forth. Adjust the lighting area to allow easy viewing.
6
Attaching filters [C-DSLS]
Rotate and remove the filter cap at the end of
the lamp house and insert a filter. Use filters
that have a diameter of 33 mm and a width of
1.5 ~ 2 mm.
2
P-ICI2 Coaxial Episcopic Illuminator
Magnification of the P-ICI2 Coaxial
Episcopic Illuminator is 1.5X. Total
magnification is obtained by multiplying 1.5
by the total magnification of the basic set of
the microscope.
Note) Usable magnification for the P-ICI2
Coaxial Episcopic Illuminator is
restricted. Use the magnification
within the range shown in the right
Filter
ø33 mm
Filter cap
Objective
Zooming
magnification
Acro 0.5X 3X or higher
Plan 1X 1.5X or higher
Plan 1X ERGO 1.5X or higher
Plan Apo 0.5X 4X or higher
Plan Apo 1X 2X or higher
ED Plan 1.5X 2X or higher
ED Plan 2X Not usable
HR Plan Apo 0.5X 2.5X or higher
HR Plan Apo 1X 2.5X or higher
HR Plan Apo 1.6X Not usable
Fiber illuminator
Light guide
Luminous section
Main body
Eyepiece tube
clamp screw
Incident section
Light guide clamp screw
Light guide mount
Fiber illuminator (C-FI115 for
100/120V region and C-FI230
for 230V region) available on
order is necessary for the
light source.
Read also the Instruction
Manual of the Fiber
Illuminator C-FI115/C-FI230.
Illuminator