Instruction Manual: AQUA kit for MA-3 Solo NIC-TD-0000187-01
Nippon Instruments Corporation 11
5. MAINTENANCE
DAILY INSPECTION
Measures Against
Abnormality Etc.
Inspection
Before Start
of Work
There shall be no dirt or wetting.
The tube shall not be disconnected.
A warning or an error shall not
occur at a startup.
See the instruction manual of
the MA-3 Solo.
There shall be no cloudiness.
Disposal of reagent, cleaning
of container
There shall be no crack or chip.
Waterdrops shall not stick.
Checking at
Time of
Measurement
Bubbling shall be carried out.
Checking of tube connection.
Checking of impinger
connection.
The correlation coefficient (r) shall
be 0.999 or more.
Preparing new reagents.
Replacement of mercury
collector tube.
RECOMMENDED REPLACEMENT COMPONENTS
We recommend that the following components be replaced periodically to use the product
stably, though their lifetimes vary depending on the usage condition:
* The recommended replacement period is not the guaranteed lifetime.
ERRORS AND WARNINGS
These messages are the same as that of the MA-3 Solo main unit. For details, see the
instruction manual of the MA-3 Solo main unit.
When waterdrops have stuck
When the diffuser tube has been clogged.
Impinger outside
cylinder
When it is impossible to remove white dirt
When the sensitivity has lowered or high-
concentration contamination has occurred