26
System Out of Range Warning Messages
CAUTION!
TRUEtrack measures blood glucose
results from 20-600 mg/dL.
If blood glucose result is less than
20 mg/dL, “Lo” appears in Meter
Display.
If blood glucose result is greater
than 600 mg/dL, “Hi” appears in
Meter Display.
Always repeat test to confirm Low (“Lo”) and
High (“Hi”) results. If results still display “Lo
orHi, call your Doctor or Diabetes Healthcare
Professional immediately.
Note:Lo” results are included in the Morning
Average as 20 mg/dL. “Hi” results are
included as 600 mg/dL.