36
Messages (continued)
Display
Reason
Meter Error
Test Strip Error,
Very high blood
glucose result
(higher than
600 mg/dL)
Test Strip
Removed
During Test
Meter
Error
Code Chip
Error
Action
Call for
assistance.
Retest with new Strip. If error
persists, call for assistance.
If you have symptoms such
as fatigue, excess urination,
thirst, or blurry vision, follow
your healthcare professional’s
advice for high blood glucose.
Retest with new
Test Strip. If error
persists, call for
assistance.
Call for
assistance.
Remove Code
Chip and reinsert.
If error persists,
call for assistance.