31
The following section explains how to run a Quality Control
Test with one of the 2 StatStrip Lactate Control Solutions.
ReadtheStatStripLactateSolutionpackageinsert
sheetforcompleteinstructions,indications,precau-
tions,andlimitationsofthesystem.
1. Insert a test strip into the meter. All segments of the
screen will display for 2 seconds. Then a ashing blood
drop will display.
Testing a Quality Control Solution
NOTE:Ifstripisremoved
beforetheteststarts
orisnotusedfor
over2minutes,the
screenwillgoblank.