13
ProCair Plus User Guide
Operation - Control Panel
G Alarm Mute
Turns audible alarm off temporarily. Press to mute
the alarm.
Alarm will resound in 20 minutes if the issues has
not been resolved, or immediately if new fault
detected.
H Max Infl ate
Rapidly infl ates mattress to maximum pressure
in Static mode. All pressure setting lights will be
illuminated, no other pressure settings can be
selected whilst the unit is in Max Infl ate mode.
System will automatically return to Alternating
mode after 20 minutes.
I Lock/Unlock
Lock and unlock the control unit panel to prevent
unwanted interference.
Press and hold the button for a minimum of three
seconds – a beep sounds and the light illuminates
to indicate system is locked. When locked, only
the Alarm Mute and Lock/Unlock button remain
operational.
Press again for at least three seconds to unlock
(beep sounds and light turns off).
J Alternate
Press to set Alternating mode (alternate cells
cyclically infl ating and defl ating). Light illuminates
to indicate Alternating mode is active.
K Fowler Boost
When illuminated, indicates Fowler Boost feature
is active. When mattress is inclined to 30° or more,
the system automatically activates Fowler Boost
to increase mattress pressure to accommodate
extra load in sacral region. When mattress declines
to a fl at position, Fowler Boost will automatically
deactivate.
L Static
Press to set Indefi nite Static mode (all cells infl ated
with no dynamic alternation). The pressure level
will be the level selected in previous mode and LED
of selection level will be on. User to manually adjust
the comfort level.
Indefi nite Static mode does not revert automatically
to Alternating mode.
M Power
Press and hold the Power button for at least 3
seconds to turn the system power on and off.
N Service Due
When illuminated, indicates system is due for
periodical service and maintenance procedures.
Please contact Novis Healthcare for support.
Equipment must only be serviced by qualifi ed
personnel.
O Power Failure
When illuminated, indicates no power supply to the
control unit or battery (ie battery power is depleted).
During a power failure, the control unit will
automatically switch to FailSafe mode, as indicated
by the Battery in Use indicator. Once battery power
is depleted, the Power Failure indicator fl ashes
amber with audible alarm to alert carer.
3 SEC 3 SEC
Green light
power on
Amber light
standby power, power
source connected