EasyManua.ls Logo

Novis ProCair Plus - System Preparation

Novis ProCair Plus
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
ProCair Plus User Guide
System Preparation
4
Check all internal quick release air hose
connectors are securely connected. Open
the top cover by unzipping the CPR-side of the
mattress (zipper located at foot end), check
each connector is secure by pushing the air
h ose co nn ec to rs to ge th e r ( the re sh ou ld b e
no movement). If a connection is open, a click
will be heard once connector is rmly closed.
5
Hang the control unit over the foot end of the
bed, using the inbuilt spring loaded hanging
hooks. Pull the hooks by the rubber tabs to
prevent accidentally trapping your ngers.
E nsu re i t i s s ec u re b efo re u se ; f ai lu re to do s o
could result in equipment damage.
6
Connect the umbilical cord to the side of
the control unit. Listen for two clicks as
con rmation the connector is locked in place.
Straighten any twists in the umbilical cord air hoses
to ensure uninterrupted air ow between the control
unit and mattress.
Ensure the umbilical cord is not trapped between the
mattress and bed. Failure to do so could result in an
under in ated mattress leading to patient injury.

Related product manuals