25
ProCair Plus User Guide
Technical Specifi cations
Guidance and Manufacturer’s declaration-electromagnetic
emissions
The ProCair Plus control unit is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the ProCair Plus control unit should ensure that it is used in such an environment.
Guidance and Manufacturer’s declaration-electromagnetic
immunity
The ProCair Plus control unit is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the ProCair Plus control unit should ensure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions CISPR 11 Group 1
The ProCair Plus control unit uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B
The ProCair Plus control unit is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fl uctuations /fl icker
emissions IEC 61000-3-3
Compliance
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Electrostatic
discharge(ESD) IEC
61000-4-2
Contact ±8 kV Contact ±8 kV
F lo o rs sho uld b e w oo d , c on cre te or c e ram ic
tile. If fl oors are covered with synthetic
material, the relative humidity should be at
least 30%
Air ±2, ±4, ±8, ±15 kV Air ±2, ±4, ±8, ±15 kV
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply lines
± 2kV for power
supply lines
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
± 1kV for input/
output lines
Not applicable
Surge
IEC 61000-4-5
± 0.5, ±1 kV line(s) to
line(s)
± 0.5, ±1 kV line(s) to
line(s)
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
± 0.5, ±1, ± 2kV line(s) to
earth
Not applicable