ProCair Plus User Guide
Technical Speci cations
Manufacturers declaration-electromagnetic immunity Test
speci cations for Enclosure Port Immunity to RF wireless
communications equipment
The control unit is intended for use in the electromagnetic environment (for home healthcare and professional
healthcare) speci ed below. The customer or the user of the control unit should assure that it is used in such an
environment.
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
A For some services, only the uplink frequencies are included.
B The carrier shall be modulated using a 50 % duty cycle square wave signal.
C As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent
actual modulation, it would be worst case.
Test frequency
(MHz)
Band
A
(MHz) Service
A
Modulation
B
Maximum
power (W)
Distance (m)
Immunity test
level (V/m)
Compliance level
(V/m) (for home
healthcare)
385 380390 TETRA 400
Pulse
modulation
B
18 Hz
1.8 0.3 27 27
450 430-470
GMRS 460, FRS
460
FM
C
±5 kHz
deviation 1
kHz sine
2 0.3 28 28
710
704-787 LTE Band 13, 17
Pulse
modulation
B
217 Hz
0.2 0.3 9 9
74 5
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820, CDMA
850, LTE Band 5
Pulse
modulation
B
18 Hz
2 0.3 28 28
870
930
1,720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4, 25; UMTS
Pulse
modulation
B
217 Hz
2 0.3 28 28
1,845
1,970
2,450 2 400-2 570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE
Band 7
Pulse
modulation
B
217 Hz
2 0.3 28 28
5,240
5 100-5 800
WLAN
802.11a/n
Pulse
modulation
B
217 Hz
0.2 0.3 9 95,500
5,785