7
ProCair Plus User Guide
Safety Precautions
Protection Against Hazards
Fluids
Avoid spilling uids on any part of the control unit. If
spills do occur:
Turn off control unit power and disconnect
the unit from mains electricity supply.
Immediately clean fluids from the casing by
wiping with a soft cloth.
Ensure there is no moisture in or near the power
inlet, control handset and power cord before
reconnecting the power supply.
Check the operation of controls and other
components around the spill area.
Fluid or liquid remaining on the electronic controls
can cause corrosion that may cause the electronic
components to fail. Component failures may cause
the unit to operate erratically, possibly producing
potential hazards to patient and carers.
Explosion Hazard
Equipment is not suitable for use in the presence of
a ammable anaesthetic mixture with air, oxygen or
nitrous oxide.
Do not use in the presence of smoking
materials or open flame – air flowing
through the mattress will support
combustion.
Do not open the control unit – risk of
electrical shock. Refer servicing to qualified
service personnel.
Disposal
Dispose of all components (control unit including
batteries, air lter, air cells, mattress cover
and base) according to local procedures and
regulations or contact Novis Healthcare for advice.
Power Cord
The system should never be operated with a worn
or damaged power cord. Keep the cord away from
heated surfaces. Should the power cord be found to
be worn or damaged, contact Novis Healthcare for
a replacement.
Interference
Although this equipment conforms to the
intent of directive IEC 60601-1-2
1
in relation to
Electromagnetic Compatibility, all electrical
equipment may produce interference. If
interference is suspected, move equipment away
from sensitive devices or contact Novis Healthcare.
1 IEC 60601-1-2. Medical Electrical Equipment - Part
1: General Equipments for Safety, Amendment No. 2.
Collateral Standard. Electromagnetic Compatibility
Requirements and Test).