Fig. 5
Notice
Use only clean water to maintain the Clean Head System.
Half fill a cup with clean water. Operate the handpiece and, while it is
running, immerse half of the handpiece head in the water, (Fig. 5)
Intermittently stop and start the handpiece in the water for about 5
seconds. While still running, remove the handpiece from the water.
Stop the handpiece and wipe it dry. If debris remains, repeat the
process until the area is clean.