    zacharyperkinsJul 27, 2025
    How to fix motor rotation failure error in NSK Medical Equipment?
    • J
      John RobinsonJul 27, 2025
      A 'Motor Rotation Failure Error' can occur due to handpiece attachment failure or motor failure. Check that the chuck is securely closed. If the error continues, request repair.
  • A
    andrewmedinaJul 31, 2025
    How to resolve a system error on NSK Medical Equipment?
    • B
      baileymicheleJul 31, 2025
      A system error may be caused by a memory failure, erroneous memory, or connecting a USB flash drive other than the type stipulated (mouse, printer, etc.). Request a repair to resolve this issue.
  • M
    Morgan PatelAug 4, 2025
    What to do if NSK Medical Equipment shows a breaking error?
    • A
      Alex HartAug 4, 2025
      A 'Breaking Error' can occur due to abnormal voltage or failure in the start/stop switch circuit. If rotation and stop are repeated too quickly, a circuit limiting acceleration might activate. Wait a few seconds before using again. If the error persists, request a repair.
  • N
    Nicholas SnyderAug 9, 2025
    What causes motor sensor error in NSK Medical Equipment?
    • J
      John MitchellAug 9, 2025
      A motor sensor error can be caused by motor sensor failure (Hall IC), motor cord failure, or water ingress into the motor. If water has entered the motor, make sure to put a Protection Plug when Thermo-disinfecting it. Request repair.
  • M
    Matthew CurtisAug 13, 2025
    Why am I getting a pump excess voltage error on my NSK Medical Equipment?
    • J
      Juan LewisAug 13, 2025
      An excess voltage error with the pump can be due to an incorrectly fitted irrigation tube loading the pump roller, or pump failure. Check the irrigation tube. If the error persists even with a normal irrigation tube, request repair.
  • C
    Cheryl ClarkAug 17, 2025
    What to do if excessive current is detected in NSK Medical Equipment?
    • D
      David RussellAug 17, 2025
      If you see an 'Excessive Current Detected' error, it could be due to extended use under heavy load or a short circuit in the motor. Ensure the Motor Cord is securely re-connected, as insufficient electrical contact may be the cause. If the error persists, request a repair.
  • A
    Amanda MartinAug 21, 2025
    Why does the NSK Medical Equipment control unit overheat?
    • M
      Michael BenjaminAug 21, 2025
      The control unit might be overheating due to extended use under heavy load or operation in an extremely high-temperature environment. Allow the unit to cool down before resuming use. Ensure the periphery of the main Control Unit is well-ventilated. If the error persists, request repair.

Summary

Safety Precautions

Control Unit and Foot Control

Control Unit Key Functions

Explains the function of each key on the control unit.

Control Unit LCD Display Overview

Details the information displayed on the control unit's LCD screen.

Installation

Setting

Operation

Micromotor Operation Programming

Memorizing and setting up 8 programs for micromotor operation.

Implant System Configuration

Configuring different implant systems and their parameters.

Handpiece Calibration Procedure

Calibrating the handpiece for accurate torque and resistance recognition.

Standard Operation Guide

Step-by-step guide for normal operation of the system.

USB Data Storage and Management

Storing and managing treatment data on a USB flash drive.

Protection Circuit Operation

Error Codes and Troubleshooting

Equipment Maintenance

Sterilization Procedures

NSK Surgic Pro Specifications

General IconGeneral
Motor SpeedUp to 40, 000 RPM
DisplayLCD
Power SupplyAC power supply
Speed Range200~40, 000 min-1
Rotation SpeedUp to 40, 000 RPM
Head DiameterVaries depending on the handpiece
Light SourceLED
AutoclavableMotor, Handpiece

Related product manuals