  • J
    Joseph ThompsonAug 14, 2025
    What to do if NSK Surgic Pro2 shows error E1?
    • S
      Steven BeckerAug 14, 2025
      Error E1 on your NSK dental equipment can be caused by extended use under heavy load, a short circuit in the motor, or a motor cord failure. First, ensure the motor cord is securely re-connected, as insufficient electrical contact may be the issue. If the error persists after re-connecting the cord, request a repair.
  • T
    Troy Brown IIAug 17, 2025
    What to do if NSK Surgic Pro2 Dental equipment shows error E2?
    • M
      Matthew MolinaAug 17, 2025
      Error E2 on your NSK dental equipment indicates insufficient contact of the power cord or a failure of the power supply circuit. Ensure the power cord is securely re-connected. If the error continues, request a repair.
  • D
    Duane DavidsonAug 21, 2025
    What to do if NSK Surgic Pro2 Dental equipment shows error E5?
    • A
      Amanda RichardsonAug 21, 2025
      Error E5 on your NSK dental equipment is due to abnormal voltage generated in the start/stop switch circuit or a failure in that circuit. If rotation and stop are repeated quickly, a circuit that suppresses sudden acceleration may activate. Wait a few seconds before using the equipment again. If the error persists, request a repair.
  • D
    dominiquemooreAug 25, 2025
    What to do if NSK Surgic Pro2 Dental equipment shows error E6?
    • A
      Anthony LeachAug 25, 2025
      Error E6 on your NSK dental equipment may be due to a bur or drill attachment failure, handpiece failure, or motor failure. Check that the chuck is securely closed. If the error persists, request a repair.
  • O
    obrownAug 27, 2025
    What to do if NSK Surgic Pro2 Dental equipment shows error E3?
    • J
      James FloresAug 27, 2025
      Error E3 on your NSK dental equipment can be triggered by a motor sensor failure, an unplugged motor cord, a motor cord (signal line) failure, or water ingress into the motor. Make sure the motor cord is securely re-connected. Note that electrical contact may be insufficient. If the error persists, request a repair. Also, remember to attach a protection plug when cleaning in a washer-disinfector.
  • B
    Brian SanchezAug 31, 2025
    Why NSK Surgic Pro2 Dental equipment shows error E4?
    • M
      mark32Aug 31, 2025
      Error E4 on your NSK dental equipment can be caused by overheating from extended use under heavy load, or using the control unit immediately after it has been exposed to high temperatures. Allow the unit to cool down before use, ensuring the periphery of the control unit is well-ventilated. If the error continues, request a repair.
  • L
    Lonnie StevensonSep 3, 2025
    What to do if NSK Dental equipment shows error E7?
    • A
      Ashlee MosesSep 3, 2025
      Error E7 on your NSK dental equipment may be due to an incorrectly fitted irrigation tube caught on the pump roller, or an irrigation pump failure. Check the irrigation tube. If the error continues, request a repair.
  • J
    Jonathan ColeSep 7, 2025
    What to do if NSK Surgic Pro2 Dental equipment shows error E0?
    • K
      kimberly65Sep 7, 2025
      Error E0 on your NSK dental equipment is a self-check error. Request repair.
  • D
    Denise RaySep 10, 2025
    What to do if NSK Surgic Pro2 shows error E8?
    • A
      aacostaSep 10, 2025
      Error E8 on your NSK dental equipment indicates a failure of the Bluetooth communication circuit. Request repair.

Summary

User and Intended Purpose

Target User and Application

Defines the intended user and purpose of the Surgic Pro2 system.

Safety Precautions for Handling and Operation

Critical Safety Warnings

Lists critical safety warnings regarding electric shock, electromagnetic interference, and hazardous substances.

Operational Handling Cautions

Provides cautions regarding handpiece compatibility, system operation, and EMC information.

Handling and Operation Cautions

Details precautions for handling, operation, and potential hazards like overheating and eye injury.

Service and Support Information

Mentions service requirements, spare parts availability, and required professional training.

Product Description and Components

Package Contents Overview

Lists all items included in the product package with part names and quantities.

Control Unit and Foot Control Components

Details the physical components of the control unit and foot control, including panel and display.

Foot Control Operation Details

Explains the functions of the foot control buttons and speed control pedal.

Control Panel Key Functions

Identifies and describes the keys, display elements, and their functions on the control panel.

Preparation for Use

AC Power Cord Connection

Provides instructions for connecting the AC power cord to the control unit.

Foot Control Battery Installation

Details the process of installing AAA batteries into the foot control.

Irrigation Tube Installation

Provides step-by-step instructions for installing the irrigation tube.

Motor Cord Connection Procedure

Explains how to connect the motor cord to the control unit.

Handpiece Connection Guide

Guides on how to connect the handpiece to the motor.

Pre-Treatment Operational Checks

Lists essential checks to perform before using the product on a patient.

Spray Nozzle and Drill Compatibility Check

Details how to check compatibility of internal spray nozzles with drills.

Operation Procedures

Calibration Function Overview

Explains the purpose of the calibration function and its importance for accuracy.

Conducting Torque and Speed Calibration

Provides instructions for performing torque and speed calibration with and without load.

Starting System Operation

Guides on how to start the system, select programs, and activate the motor.

Sleep Mode and Protection Circuit

Explains sleep mode and the motor overload protection circuit features.

Connectivity Features (Bluetooth, Osseo, Link Module)

Explains Bluetooth linking to Osseo 100+, VarioSurg 3, and external terminals.

Post-use Maintenance

Maintenance Preparation and Cleaning

Lists steps for preparing the unit before maintenance and manual cleaning procedures.

Sterilization Procedures

Explains the sterilization procedures for the sterilizable equipment according to EN ISO standards.

Product Storage Guidelines

Guides on proper storage of the product in a clean and suitable environment.

Routine Maintenance and Calibration

O-ring Replacement and Periodical Checks

Instructions for replacing O-rings and a checklist for periodic maintenance.

Foot Control Calibration Process

Explains how to perform calibration for the foot control pedal and buttons for optimal response.

System Setting and Configuration

Using the Setting Key for Configuration

Guides on using the setting key to change control unit and foot control operational settings.

Foot Control Pairing and Button Assignment

Covers foot control pairing and how to assign functions to its buttons.

Implant System Setup and Operation Programming

Explains implant system setup and programming system operations for procedures.

Troubleshooting Common Issues

Error Code Reference and Actions

Lists error codes, their causes, and recommended actions for troubleshooting.

Common Problem Diagnosis and Solutions

Provides solutions for common problems like blank screen, incorrect display, or connectivity errors.

Handpieces, Spare Parts, and Optional Parts

Recommended Handpieces and Spare Parts

Lists recommended handpieces and a comprehensive spare parts list with catalog numbers.

Optional Parts and Compatible Terminals

Lists optional accessories and specifies compatible terminal devices for the SurgicPro2 App.

Technical Specifications

Product Specifications Overview

Details the technical specifications for the control unit, micromotor, and foot control.

Factory Settings for Implant Systems

Lists the initial factory settings for different implant systems and programs.

EMC Information and Compliance

Electromagnetic Emissions and Immunity

Declares electromagnetic emissions, immunity, and guidance for the operating environment.

RF Equipment Separation Distances

Provides recommended separation distances for RF equipment to prevent interference.