Specifications
English
11 – 8 Factory settings (Implant system)
The following are the initial factory settings� These are only reference setting conditions, so change them to the
conditions for the implant system you use�
Implant
system
number
Program
number
Maximum
rotation speed
[min
-1
]
Torque upper
limit
[N•cm]
Rotation
direction
[F/R]
Gear ratio
[X:X]
Coolant
solution flow
level
[0-5]
Light level
[H/L]
1
1 5 2000 1200 – – F F
20:1
3 3
H
2 6 2000 25 – 45 F F 3 3
3 7 2000 25 – 50 F R 3 3
4 8 1600 25 – 20 F F 3 0
2
1 5 2000 800 – – F F
20:1
3 3
H
2 6 800 25 – 45 F F 3 3
3 7 800 25 – 50 F R 3 3
4 8 800 25 – 20 F F 3 0
3
1 5 800 400 – – F F
20:1
3 3
H
2 6 600 15 – 35 F F 3 3
3 7 500 15 – 40 F R 3 3
4 8 400 15 – 35 F F 3 0
4
1 5 800 300 – – F F
20:1
3 3
H
2 6 600 15 – 35 F F 3 3
3 7 500 15 – 40 F R 3 3
4 8 400 15 – 35 F F 3 0
5
1 5 1400 15 – 35 F F
20:1
3 3
H
2 6 1400 15 – 35 F F 3 3
3 7 1400 15 – 40 F R 3 3
4 8 800 15 – 35 F F 3 0
6
1 5 1400 1400 – – F F
20:1
3 3
H
2 6 1400 1200 – – F F 3 3
3 7 1400 1000 – – F F 3 3
4 8 1400 25 – 35 F F 3 0