Maintenance
English
7 – 3 Foot control calibration
1
10 63, 6, 8
52, 4
98
Foot control pedal and button response may become poor due to wear of parts over time� In that case, perform
calibration as described below
1
Press and hold for 2 seconds or more to switch
to setting mode�
2
Press to select “FC SET”�
3
Press
4
Press to select “FC CAL”�
5
Press to select “ON”�
6
Press and confirm that “---:-” is displayed�
7
Fully depress the pedal and all of the buttons on the
foot control�
8
When “000:0” is displayed, calibration is successful�
Press
9
Make sure that “OK” is displayed on the LCD display�
10
Press twice to end setting mode�
CAUTION
Do not press the enter key when other than “000:0” is displayed�
If calibration is not performed correctly, “NG” is displayed on the LCD
display� In this case, start again from step 1�