37
Setting
English
<Setting values of foot control button>
Setting
value
Function
Press and release button Press and hold button more than 2 seconds
0 Invalid Invalid
1 Coolant solution flow level selection (up) Switching operation between this product and
linked device
2 Program number selection (up) Program number selection (down)
3 Direction of rotation switching Motor LED illumination while button is depressed
(Motor will not rotate)
4 Rotation speed setting (up) Rotation speed setting (down)
5 Rotation speed setting (up) Rotation speed setting (fast forward to up-direction)
6 Rotation speed setting (down) Rotation speed setting (fast forward to down-direction)
7 Torque setting (up) Torque setting (down)
8 Torque setting (up) Torque setting (fast forward to up-direction)
9 Torque setting (down) Torque setting (fast forward to down-direction)
A Coolant solution flow level selection (up) Coolant solution flow level selection (0:no coolant
solution flow)
B Program number selection (up) Implant system selection (up)
C Implant system selection (up) Implant system selection (down)
D Motor LED brightness selection
E Motor LED illumination while button is depressed (Motor will not rotate)
F Irrigation while button is depressed (Motor will not rotate)
8 – 2 Implant system
Eight programs can be set for each of the eight different implant systems�
When you set the program, refer to values the implant manufacturer recommends�
How to change implant system
Press the and on simultaneously to change the implant system�
* Factory settings: See “11 – 8 Factory settings (Implant system)”�