12 EMC Information (Electromagnetic
Compatibility Information)
Guidance and manufacturer’s declaration - Electromagnetic Emissions.
The product is intended for use in the electromagnetic environment specified below� The customer or the user of the product should assure that is
used in such an environment� Guidance and manufacturer’s declaration electromagnetic emissions�
RF emissions Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11/EN 55011
Group 1 The product uses RF energy only for its internal functions�
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment�
Harmonic emissions
CISPR 11/EN 55011
Class B The product is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes�
Emissions test
IEC 61000-3-2/EN 61000-3-2
Class A
(except 120V)
Voltage fluctuations/flicker emissions
IEC 61000-3-3/EN 61000-3-3
Complies
(except 120V)
Guidance and manufacturer’s declaration - Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below�
The customer or the user of the product should assure that it is used in such an environment�
Immunity test IEC/EN 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2/EN 61000-4-2
± 8kV contact
± (2,4,8)15kV air
± 8kV contact
± (2,4,8)15kV air
Floors should be wood, concrete or ceramic
tile� If floors are covered with synthetic
material, the relative humidity should be at
least 30%�
Electrical fast transient/burst
IEC 61000-4-4/EN 61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
± 2kV for power supply lines
± 1kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment�
Surge
IEC 61000-4-5/EN 61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
± 1kV line(s) to line(s)
± 2kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment�
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11/EN 61000-4-11
0% UT 0�5 cycle
@ 0°,45°,90°,135°,180°,
225°,270° & 315°
0% UT 1 cycle
and
70% Ut / 25 ( 50Hz ) cycle
30 ( 60Hz ) cycle
0% Ut / 250 ( 50Hz ) cycle
300 ( 60Hz ) cycle
0% UT 0�5 cycle
@ 0°,45°,90°,135°,180°,
225°,270° & 315°
0% UT 1 cycle
and
70% Ut / 25 ( 50Hz ) cycle
30 ( 60Hz ) cycle
0% Ut / 250 ( 50Hz ) cycle
300 ( 60Hz ) cycle
Mains power quality should be that of a typical
commercial or hospital environment� If the
user of the product requires continued
operation during power mains interruptions, it
is recommended that the product be powered
from an uninterruptible power supply or a
battery�
Power frequency (50/60Hz)
magnetic Field
IEC 61000-4-8/EN 61000-4-8
30A/m 30A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment�
NOTE ‘Ut’ is the AC mains voltage prior to application of the test level�